1. The system will be capable of measuring the dispensed reagent in at least four locations across the surface of the tray.
2. The system will be able to measure dispensed volumes to an accuracy of 0.01g.
3. The system will fit the dimensions of existing trays.
4. The system will be capable of wirelessly transmitting the data to the software running host computer.
5. The system software will be capable of calculating the calibration values.
6. The system software will be able to update the calibration data to be sent to the instrument.
7. The system will expect to work inside the instrument, without any adjustments to the instrument.
8. The system may include manual steps to initiate the measurement data transfer for further processing by the system software.
2. The system will be able to measure dispensed volumes to an accuracy of 0.01g.
3. The system will fit the dimensions of existing trays.
4. The system will be capable of wirelessly transmitting the data to the software running host computer.
5. The system software will be capable of calculating the calibration values.
6. The system software will be able to update the calibration data to be sent to the instrument.
7. The system will expect to work inside the instrument, without any adjustments to the instrument.
8. The system may include manual steps to initiate the measurement data transfer for further processing by the system software.