Pathology Lab Barcode Reader Test Bench

Project number
26020
Organization
Roche Tissue Diagnostics
Offering
ENGR498-F2025-S2026
The stand shall accept standard slides
The stand should recreate 5 of the slide drawers as the parent assembly.
The stand should be able to index across all 5 slide drawers continuously
The stand shall have a means to report out the quality status of each slide.
The stand shall be able read all 5 barcode types.
The stand shall be able to read a barcode on the slide label and report back appropriate barcode information to the system.
The stand shall hold all slides in the same orientation.
The stand shall not break more than 1 in 100 slides with 95% confidence.
The size of the stand should be able to be placed on a desk or a bench top
The stand should weigh less than 40 lbs
The stand shall use the same drawer components as the parent instrument assembly (Provided by Roche)
The module shall utilize a standard 120V 60 hz power plug
The module shall not require use of external compressed air, vacuum, or fluidics.
The stand shall have mounting interface dimensions compatible with the parent instrument (specifications provided by RTD).
The system shall be able to complete a full transfer and imaging of 25 slides within 420 seconds.